Glove leak detector EX1000
Performance features
1. Compact structure, light weight, and easy to operate;
2. Wireless online detection, no need to remove gloves, easy to use;
3. Built-in lithium battery power supply, requiring no external power supply;
4. Built-in diaphragm air pump to inflate the glove, to ensure that the gas is clean, while requiring no external air source;
5. Pressure detection range up to 3000Pa, covering all glove detection pressure;
6. Supporting offline detection, which requires a glove detection bracket;
7. Fully automatic monitoring of the pressure inside the inflatable seal and gloves during the test;
8. Equipped with a variety of test formula settings to meet the needs of a variety of glove testing;
9. Multiple glove leak detectors connected, to achieve group control testing.
Leak detection principle: Pressure decay method (pressure drop method)
Meet the requirements of FDA and GMP regulations:
1. Comply with the standards: ISO14644-7.2004 and GB/T25915.7-2010;
2. Support three-level user management;
3. Support audit trail function and 21 CFR Port 11 compliant electronic recording;
4. The detection results can be printed instantly via Bluetooth, and printed after the test is completed;
5. The detection results can be exported in PDF format via USB interface, and the stored data cannot be changed.
Hydrogen Peroxide Sterilizer